TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090, SC-035-065, SC-035-135, SC-014-150, SC-018-150, SC-035-090, and SC-035-150. Intended to guide and support a guide wire during access of the vasculature.

This recall has been terminated (originally issued December 23, 2009).

Recall Initiated
November 6, 2009
Posted
December 23, 2009
Terminated
January 5, 2011
Recall Number
Z-0497-2010
Quantity
70
Firm Location
Plymouth, MN
Official Source
View on FDA website ↗

Reason for Recall

Catheter may crack near the radiopaque marker band at the distal end of the device during use. Cracking of the catheter shaft in this location can result in material separation, and potential embolization.

Distribution

Nationwide Distribution -- USA, including states of CA, FL, GA, IL, IN, IA, KS, KY, MI, MT, NJ, PA, SD, TN, TX, and WI.

Lot / Code Info

Lot #: 7828282, 7792290, 7792584, 7805570, 7805797, 7806392, 7820252, 7790666, 7791887, 7803307, 7835331, 7822400, 7822593, 7800446, 7800555, 7800756, 7800779, 7800809, 7801822, 7801875, 7803305, 7803306, 7820273, 7834779, 7834845, 7824905, and 7832205.

Root Cause

Component design/selection

Action Taken

Consignees were sent a Medical Device Recall letter, dated 11/6/09. The letter was addressed to Risk Manger or Cath Lab Manage. The letter provided the Issue Summary, identified the Affected Product, and stated the Required Action, which asked consignees to locate and remove the affected product. Questions should be directed to ev3 Customer Service at 800-716-6700.