Heartstart FR2+ automated external defibrillator models M3861A and M3841A are NOT equipped with the ECG Module. Model M3861A is under the Philips label and M3841A is under the Laerdal label.

This recall has been terminated (originally issued December 8, 2009).

Company
Philips Medical Systems
Recall Initiated
October 1, 2009
Posted
December 8, 2009
Terminated
September 27, 2013
Recall Number
Z-0062-2010
Quantity
see Z-0063-2010
Firm Location
Seattle, WA
Official Source
View on FDA website ↗

Reason for Recall

A potential exists for a component failure, which, if it occurs, could result in the inability to deliver therapy.

Distribution

Domestic and Worldwide distribution to: United States, Canada, Australia, Austria, Bulgaria, China, Finland, France, Germany, Gibraltar, Great Britain, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lebanon, Malaysia, Mexico, Netherlands, Norway, Portugal, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Arab Emirates.

Lot / Code Info

The following letter codes are used in conjunction with the catalog numbers to indicate language of Instructions for Use (IFU) and FR2+ voice prompts: ABA=U.S. English; ABU=U.K. English; ABD=German; ABE=Spanish; ABF=French; ABH=Dutch; ABI=Hong Kong; ABJ=Japanese; ABN=Norwegian; ABO=Taiwanese; ABS=Swedish; ABX=Finland; ABZ=Italian; AB2=China; and  AC4=Portugal.  Additional (non-language related) letter codes for product: A01=Product with TSO (Technical Standard Order) approved battery; and RF=Service Replacement.

Root Cause

Nonconforming Material/Component

Action Taken

On September 28, 2009 Philips issued a nationwide press release. On October 1, 2009 the firm issued notification letters to consignees arranging for the return and replacement of the affected units. The firm also set up a page on their website with a serial number look-up tool. For further information or support concerning this issue, customers are asked to contact Philips at 1.800.263.3342 or visit www.philips.com/FR2PlusAction for the serial number look-up list, answers to frequently asked questions, directions for the battery insertion test, and the replacement AED selection form.