SILHOUETTE XTRAFLO DEVICE WITH SL-6 HYDROPHILIC COATING, MODEL NUMBER: B3857, SIZE 6F x 26cm Urological stents are disposable single-use temporary indwelling catheters that allow urine to drain through obstructed or strictured ureters.

This recall has been terminated (originally issued September 16, 2008).

Company
Applied Medical Resources Corp 22872 Avenida Empresa # 3 Rancho Santa
Recall Initiated
May 20, 2008
Posted
September 16, 2008
Terminated
April 14, 2010
Recall Number
Z-2220-2008
Quantity
4
Firm Location
Margarita, CA
Official Source
View on FDA website ↗

Reason for Recall

The product is being recalled due to possible movement or dislodgement of the positioner marker band during use.

Distribution

Nationwide to: AL, AZ, CA, CO, DC, FL, GA, KY, IL, LA, MI, MN, NC, NJ, NY, PA, TN, TX, WA, WI and Worldwide to: Australia & South Africa

Lot / Code Info

LOT NUMBER: 1059245

Root Cause

Mixed-up of materials/components

Action Taken

On May 20, 2008, Applied Medical began distribution of an Recall Notice with fax-back attachment ( Recall Notification Confirmation Form ) to all customers in the United States via Applied's Direct Sales Force. The Notice informed customers of the problem that caused the initiation of the recall and requested they complete and fax-back the Recall Notification Confirmation Form . The letter also instructed their customers to remove and return affected product to Applied Medical immediately via FedEx. If the customers had an questions they were to contact Mr. Frans VandenBroek at (949) 7l3-8369 or by e-mail at fvandenbroekappliedmedical.com.