RecallDepth

Stella Interdental Osteotome, 4mm, 6mm and 8mm. Model numbers 01-0575, 01-0576 and 01-0578. Product is intended to be used during the interdental osteotomy (bone cutting).

This recall has been terminated (originally issued September 11, 2008).

Company
Biomet Microfixation, Inc.
Recall Initiated
April 16, 2008
Posted
September 11, 2008
Terminated
September 11, 2009
Recall Number
Z-1871-2008
Quantity
36
Firm Location
Jacksonville, FL
Official Source
View on FDA website ↗

Reason for Recall

The working tips of the Stella Osteotomes are thicker than specification.

Distribution

Distributed domestically to Missouri, New Mexico and Texas, and countries of Canada and Germany.

Lot / Code Info

Lot numbers: 381250, 381251, 439970, 381240, 356510,513250 and 354320.

Root Cause

Nonconforming Material/Component

Action Taken

The firm notified its consignees on 4/16/08 by certified mail with return receipt. Sub-accounts were contacted by email. All returned product will be stored in quarantine until a final determination of disposition is made.