Infinity Docking Station (IDS) (transportable patient monitor) Repair Kit, part number 7262814

This recall has been terminated (originally issued December 19, 2006).

Company
Draeger Medical, Inc.
Recall Initiated
November 17, 2006
Posted
December 19, 2006
Terminated
November 19, 2008
Recall Number
Z-0303-2007
Quantity
23,345 units - 4336 US
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.

Distribution

Nationwide

Lot / Code Info

Part number 7262814

Root Cause

Other

Action Taken

The recaling firm issued a recall letter to the US customers 11/17/2006 via certified mail. An authorized service representative will inspect each potentially affected unit. If a unit is found to have an affected docking station, the unit will be repaired or replaced free or charge. Notification to international customers will occur through the international subsidiaries.