Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.

This recall has been terminated (originally issued November 14, 2006).

Company
Baxter Healthcare Renal Div
Recall Initiated
September 6, 2006
Posted
November 14, 2006
Terminated
December 20, 2006
Recall Number
Z-0169-2007
Quantity
2,823 units
Firm Location
Mc Gaw Park, IL
Official Source
View on FDA website ↗

Reason for Recall

The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring.

Distribution

Nationwide and Hong Kong, Korea and China

Lot / Code Info

all serial numbers

Root Cause

Other

Action Taken

Urgent Device Correction letters dated 9/6/06 were sent to the Meridian accounts on the same date via UPS second day, to the attention of the Hemodialysis Administrator. The letters informed the accounts of the potential risk associated with a post-pump bloodline kink when using pre-pump arterial pressure monitoring. The accounts were provided with a revised Meridian Operator''s Manual which includes the following warnings: 'The lower venous limit should be set as close to the currently displayed venous pressure as conditions allow.' and 'Reliance on the pressure alarms should not substitute for clinical monitoring of the patient for adverse events.' Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 2. General questions were directed to The Center for One Baxter at 1-800-422-9837.

More recalls by Baxter Healthcare Renal Div

View all recalls by this company →