Terumo Advanced Perfusion System 1 Data Interface CDI 100; Catalog number 802558.

This recall has been terminated (originally issued December 30, 2006).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
October 12, 2004
Posted
December 30, 2006
Terminated
February 1, 2007
Recall Number
Z-0327-2007
Quantity
4740 for all affected products
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.

Distribution

Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Lot / Code Info

Serial numbers 00008 through 00011, 00013, 00020 through 00030, 00032 through 00034, 00036 through 00059 and 00061 through 00078.

Root Cause

Other

Action Taken

U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.