COULTER GEN.S Series Hematology Analyzer, PN 6605632, 6605632R; GENS: 6605381, 6605381R, 6605360, 6605360R, 6605470, 6605470R

This recall has been terminated (originally issued February 4, 2006).

Company
Beckman Coulter Inc.
Recall Initiated
December 15, 2005
Posted
February 4, 2006
Terminated
January 26, 2012
Recall Number
Z-0475-06
Quantity
1134
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

There is a risk of sample misidentification when processing samples in the Manual aspiration mode.

Distribution

Nationwide and Canada

Lot / Code Info

PN 6605632, 6605632R; GENS: 6605381, 6605381R, 6605360, 6605360R, 6605470, 6605470R All Software Versions

Root Cause

Other

Action Taken

A Product Corrective Action letter will be mailed the week of Dec 19, 2005 to all COUL TER@ LH700 Series System & COUL TER@ GEN.STM System customers instructing them of a rare risk of sample misidentification. (Letter will be sent by US mail)