Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000.

Company
Perkinelmer Las Inc
Recall Initiated
September 8, 2004
Posted
September 30, 2004
Terminated
October 29, 2008
Recall Number
Z-1499-04
Quantity
49 kits
Firm Location
Norton, OH

Reason for Recall

There is an increase in the number of normal specimen results below the lowest calibrator and based on a report that the normal controls were running low, sometimes out of the specified range.

Distribution

The kits were distributed to 5 consignees, located in Indianapolis, IN; Jefferson City, MO; Austin, TX; Madison, WI; and Minneapolis, MN.

Lot / Code Info

Lot #244468

Root Cause

Other

Action Taken

The firm contacted their consignees by letter sent 9/8/04.