GE OEC, InstaTrak 3500 Plus FluoroTrak and Fluoro CAT Spinal Applications used with Regular and Super 9800 and 9600 Arm Systems. GE Health Medical Systems, Inc, Salt Lake City, UT 84116-2862

This recall has been terminated (originally issued November 10, 2007).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
March 23, 2007
Posted
November 10, 2007
Terminated
April 16, 2012
Recall Number
Z-0181-2008
Quantity
61 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Tracking Accuracy - Image intensifier supplier's change on the C-Arm could lead to patient injury, due to incorrect anatomical location of the instrument during a procedure.

Distribution

Class II Recall - Worldwide Distribution - USA including states of AZ, CA, FL, GA, IA, IL, KS, LA, MA, MD, MI, NC, NY, OK, PA, SC, TX, UT, VA, and WI, and countries of Canada, China, Germany, Hong Kong, Italy, Japan, Mexico, and Spain.

Lot / Code Info

Serial numbers: IT35052, IT35070, IT35108, IT35109, IT35129, IT35130, IT35132, IT35133, IT35134, IT35139, IT35146, IT35152, IT35166, IT35169P-RC-1, IT35173, IT35179P, IT35181, IT35185P-G-RC, IT35187P, IT35189P, IT35193P, IT35214P, IT35216P, IT35223P, IT35225P, IT35226P, IT35234P, IT35244P, IT35245P, IT35251P, IT35269P, IT35280P, IT35286P, IT35289P, IT35292P, IT35293P, IT35296P, IT35299P, IT35301P, IT35302P, IT35304P, IT35311P, IT35312P, IT35316P, IT35323P, IT35324P, IT35330P, IT35339P, IT35344P, IT35375P, IT35377P, IT35379P, IT35381P, IT35382P, IT35383P, IT35406G, IT35435G, IT35458G, IT35468G.

Root Cause

Component change control

Action Taken

Consignees were sent an Urgent Recall Notice letter, via mail dated 10/11/07, advising users to stop usage of the FluoroTrak and FluoroCAT spinal applications on the InstaTrak 3500 Plus System. Once a solution is identified, a representative will contact consignees to arrange for installation of the solution at no charge.