Rubella IgM ELISA Test Kit. It is an enzyme-linked immunoabsorbent assay designed for the qualitative detection of IgM antibodies to Rubella virus in human serum. 1 kit (96 tests). Prepared for Sigma Diagnostics, Inc., St. Louis, MO 63178. EIA514-B

Company
Zeus Scientific, Inc.
Recall Initiated
February 24, 2000
Posted
July 30, 2003
Terminated
October 25, 2004
Recall Number
Z-1065-03
Quantity
205
Firm Location
Branchburg, NJ

Reason for Recall

The high positive control/low positive standard ratio recovering too low.

Distribution

The two lots were shipped to St. Louis, MO. No subrecall was initiated.

Lot / Code Info

Product Number: 514-B (Sigma Diagnostics) Lot Numbers: 020K6491 Exp. 10-02-2001; 090K6483 Exp. 5-31-2002.

Root Cause

Other

Action Taken

Recall letters sent out via DHL on 2/24/2000