Angiotensin I [125I] Radioimmunoassay Kit, 500 Tube Catalog Number: NEA105 For in-vitro diagnostic use

Company
Perkinelmer Life Sciences, Inc.
Recall Initiated
July 28, 2003
Posted
September 12, 2003
Terminated
November 19, 2003
Recall Number
Z-1207-03
Quantity
48 Kits
Firm Location
Boston, MA

Reason for Recall

Incorrect calibration standard value may result in 25% higher assay values for controls and patient samples

Distribution

MD, NC, CA, VT, IL, MI, IL, NY, DC, MN, HI, WA, UT Nationwide. International: Belgium, Germany, Israel, UK, Canada, Hugary, Switzerland, Korea

Lot / Code Info

Lot 181611, Expiration 8/27/03

Root Cause

Other

Action Taken

PerkinElmer notified all customers who received the lots by telephone by July 23, 2003 followed by letter dated July 25, 2003, which includes instructions for documenting receipt and lot removal either by destroying or returning product to PerkinElmer