Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators specifically for pediatric patients.

This recall has been terminated (originally issued February 13, 2012).

Company
Bio-Detek, Inc.
Recall Initiated
December 30, 2011
Posted
February 13, 2012
Terminated
December 11, 2012
Recall Number
Z-1011-2012
Quantity
3,080 units
Firm Location
Pawtucket, RI
Official Source
View on FDA website ↗

Reason for Recall

Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.

Distribution

Worldwide Distribution -- US and Canada.

Lot / Code Info

Lot Numbers 0510 thru 4511

Root Cause

Process control

Action Taken

Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.