RecallDepth

The Trimble RTS873 robotic total station is a solution designed for construction professionals seeking precision, efficiency, and reliability in their layout tasks. The integrated Class 2 green aiming laser system enhances visibility and accuracy, even on uneven surfaces. It is ideal for general and specialty contractors, the RTS873 streamlines layout processes, reduces rework, and enhances overall productivity.

This recall is currently active, issued April 28, 2025. It was issued by Trimble Ab Rinkebyvagen 17 Danderyd Sweden.

Company
Trimble Ab Rinkebyvagen 17 Danderyd Sweden
Recall Initiated
March 26, 2025
Posted
April 28, 2025
Recall Number
Z-1661-2025
Quantity
127
Official Source
View on FDA website ↗

Reason for Recall

An issue has been identified in the production of the RTS873 robotic total stations that may result in the green aiming laser output power being above its laser safety Class 2 classification.

Distribution

US and OUS (South Korea and United Kingdom).

Lot / Code Info

Trimble RTS873 instruments produced from October 2023 to February 12, 2025.

Root Cause

Under Investigation by firm

Action Taken

Product Bulletin sent 03/05/2025. Customers sold affected products have been notified by Trimble and are currently in the process of contacting end customers and informing them of the up-coming service of their products.