Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.
- Recall Initiated
- June 9, 2015
- Posted
- July 31, 2015
- Recall Number
- Z-1991-2015
- Quantity
- 1047
- Firm Location
- Waltham, MA
Reason for Recall
Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).
Distribution
Nationwide Distribution.
Lot / Code Info
Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10.
Root Cause
Component change control
Action Taken
Olympus Scientific Solutions planned action: 1. Notify product managers as to potential issues 2. Distribute new table and notice to instrument owners 30 JUL 15 3. Post new table on website for access 15 JUN 15