Medline medical convenience kits packaged as: 1) POWERGLIDE DRESSING CHANGE KIT, Kit SKU DYNDH1576; 2) CENTRAL LINE TRAY W/ CHLRP SWB, Kit SKU RWM-DC57KC.

Recall Initiated
December 24, 2025
Posted
February 9, 2026
Recall Number
Z-1331-2026
Quantity
7380 kits
Firm Location
Northfield, IL

Reason for Recall

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

Distribution

US Nationwide distribution.

Lot / Code Info

1) Kit SKU DYNDH1576, UDI/DI 10193489437041, Lot Number 25EBP737; 2) Kit SKU RWM-DC57KC, UDI/DI 10889942582013, Lot Number 25EBE082.

Root Cause

Packaging change control

Action Taken

Medline issued an URGENT MEDICAL DEVICE recall notice to its consignees on 12/24/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the consignee locate and quarantine the affected units and respond to the notice using the website link: https://recalls.medline.com; Recall Reference #: R-25-257-FGX1, and the specific code listed in the notice. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who further distributed the affected its were directed to notify those to whom the kits were distributed or transferred, and include those quantities on the response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.