Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.

Recall Initiated
November 26, 2025
Posted
January 12, 2026
Recall Number
Z-1051-2026
Quantity
32 units
Firm Location
Northfield, IL

Reason for Recall

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

Distribution

US Nationwide distribution.

Lot / Code Info

Medline kit SKU PHS972096014B UDI/DI: 10889942695911 (EA) 40889942695912 (CS), Lot Number 25IBT843; Medline kit SKU CDS984289O UDI/DI: 10198459418143 (EA) 40198459418144 (CS), Lot Number 25IMA327; Medline kit SKU DYNJ903523F UDI/DI: 10198459316593 (EA) 40198459316594 (CS), Lot Number 25IMI028.

Root Cause

Nonconforming Material/Component

Action Taken

Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.