Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.
- Recall Initiated
- November 26, 2025
- Posted
- January 12, 2026
- Recall Number
- Z-1051-2026
- Quantity
- 32 units
- Firm Location
- Northfield, IL
Reason for Recall
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Distribution
US Nationwide distribution.
Lot / Code Info
Medline kit SKU PHS972096014B UDI/DI: 10889942695911 (EA) 40889942695912 (CS), Lot Number 25IBT843; Medline kit SKU CDS984289O UDI/DI: 10198459418143 (EA) 40198459418144 (CS), Lot Number 25IMA327; Medline kit SKU DYNJ903523F UDI/DI: 10198459316593 (EA) 40198459316594 (CS), Lot Number 25IMI028.
Root Cause
Nonconforming Material/Component
Action Taken
Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.