RecallDepth

Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No

This recall is currently active, issued October 29, 2025. It was issued by Medistim Asa Okernveien.

Company
Medistim Asa Okernveien
Recall Initiated
September 26, 2025
Posted
October 29, 2025
Recall Number
Z-0410-2026
Quantity
14
Official Source
View on FDA website ↗

Reason for Recall

Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

Distribution

Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.

Lot / Code Info

Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011

Root Cause

Process control

Action Taken

Firm notified impacted consignees by email on September 26, 2025. Consignees were informed that qualified service personnel will perform the correction of the device according to the standard service procedure.

More recalls by Medistim Asa Okernveien

View all recalls by this company →