Model SM-32HF-Batt; 32KW ANALOG, Mobile X-ray system

Company
Sedecal Sa Pelaya, 9 Pol. Pol. Ind. Rio De Janeiro Algete Spain
Recall Initiated
June 10, 2025
Posted
October 23, 2025
Recall Number
Z-0256-2026
Quantity
22 units

Reason for Recall

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Distribution

US: CA, IL, and NJ

Lot / Code Info

UDI/DI 08436046001466, Serial Numbers: G-33344, G29404, G29407, G30258, G30236, G63124, G60169, G26234, G26284, G25734, G38069, G-40590, G41143, G62020, G69393, G38082, G41163, G60166, G26244, G40196, G60461, G33309.

Root Cause

Device Design

Action Taken

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

More recalls by Sedecal Sa Pelaya, 9 Pol. Pol. Ind. Rio De Janeiro Algete Spain

View all recalls by this company →