BD BBL CultureSwab EZ Collection and Transport System, REF 220144, in vitro Diagnostic, SterileR, Rx Only

Company
Copan Italia Via Francesco Perotti 10
Recall Initiated
May 2, 2025
Posted
June 9, 2025
Recall Number
Z-1937-2025
Quantity
81,600 units

Reason for Recall

Swabs for specimen collection may be prone to breakage during product usage.

Distribution

US Nationwide distribution in the state of MD.

Lot / Code Info

Lot: 2415377/UDI: 38053326002145

Root Cause

Nonconforming Material/Component

Action Taken

On May 2, 2025, Copan Italia Spa issued a Urgent: Medical Device Recall" notification to affected consignees via E-Mail. Copan asked consignees to take the following steps: 1. Trace and dispose of all affected products 2. Disposal must be carried out according to local regulations. Evidence or declaration of Product disposal must be returned in writing 3. This notice should be passed on to all those who need to be aware within your organization. if you have further distributed the devices, please provide those Customers a copy of the recall notification and response form. 4. Confirm you have read and understood the recall instructions provided in the letter and submit the acknowledgment form.