Natus Brain Monitor Breakout Box, Part Number 021911; full-montage standard electroencephalograph

This recall is currently active, issued July 18, 2025. It was issued by Natus Neurology Dba Excel Tech., Ltd. (Xltek).

Company
Natus Neurology Dba Excel Tech., Ltd. (Xltek)
Recall Initiated
June 18, 2025
Posted
July 18, 2025
Recall Number
Z-2138-2025
Quantity
27 units
Firm Location
Oakville Canada
Official Source
View on FDA website ↗

Reason for Recall

Natus received two complaints of left side labeling show impedance for right side.

Distribution

US: CA, IN, MO, NC, PA, RI, TX

Lot / Code Info

UDI/DI 038283NA00153ED, Serial Numbers: 021918100008-M, 021918100014-M, 021918100015-M, 021918100017-M, 021918100019-M, 021918100020-M, 021918100021-M, 021918100025-M, 021918100026-M, 021918100028-M, 021918100032-M, 021918100034-M, 021918100036-M, 021918100038-M, 021918100039-M, 021918100040-M, 021918100041-M, 021918100044-M, 021918100046-M, 021918100047-M, 021918100048-M, 021918100049-M, 021918100050-M, 021918100051-M, 021918100052-M, 021918100054-M, 021918100055-M.

Root Cause

Process control

Action Taken

Natus issued an Urgent Medical Device Recall notice to its consignees on 06/18/2025 via USPS. The notice explained the issue, risk, and requested the user immediately cease use and quarantine pending return of the unit. A replacement device(s) will be made available to you at no cost, once you provide the completed Customer Reply Form that is listed below and return to FCA@natus.com. Technical Service will be in contact with you to arrange for a provision of replacement device(s).