FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Company
Qualigen Inc
Recall Initiated
August 26, 2024
Posted
October 8, 2024
Recall Number
Z-0022-2025
Quantity
60
Firm Location
Carlsbad, CA

Reason for Recall

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

Distribution

US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland

Lot / Code Info

UDI-DI: 00816467020099, Lot: 2405017-1, Expiration: 2025-05-15

Root Cause

Nonconforming Material/Component

Action Taken

On 8/26/24, correction notices were emailed to customers who were asked to do the following: 1) Dispose of affected TSH calibration cards. 2) Use TSH Calibration Card Lot Number 2405017-1R sent with correction notices. 3) Run TSH Quality Control samples. 4) Review and retest any affected TSH results that may have been higher than expected. 5) Notify all relevant parties within your organization or any organization that received the affected devices. 6) Complete and return response forms via email to systemsupport@qualigeninc.com If you have any questions regarding this matter, contact System Support Specialists at systemsupport@qualigeninc.com or (877) 709-2169, option 2.