Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

This recall is currently active, issued February 2, 2024. It was issued by Philips North America.

Company
Philips North America
Recall Initiated
January 6, 2022
Posted
February 2, 2024
Recall Number
Z-0974-2024
Quantity
2
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Distribution

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Lot / Code Info

UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030

Root Cause

Device Design

Action Taken

Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.