Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS

This recall is currently active, issued January 10, 2024. It was issued by Medtronic Navigation, Inc..

Company
Medtronic Navigation, Inc.
Recall Initiated
November 28, 2023
Posted
January 10, 2024
Recall Number
Z-0699-2024
Quantity
4,224
Firm Location
Lafayette, CO
Official Source
View on FDA website ↗

Reason for Recall

Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.

Distribution

Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Oman, Portugal, Spain, Switzerland, United Kingdom.

Lot / Code Info

UDI: 00613994247865 and 00643169105669(Japan)/ Serial Numbers: 2023041143, 2023051122, 2023051457, 2023051458, 2023060918, 2023051459, 2022030438, 2023041136, 2023041141

Root Cause

Device Design

Action Taken

On December 6, 2023, Medtronic Navigation, Inc. issued a Urgent: Medical Device Recall Notification via: UPS . Medtronic ask consignees to take the following actions: 1. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below. 2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form 3. Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at neuro.quality@medtronic.com. If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information. 4. This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.