Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS
This recall is currently active, issued January 10, 2024. It was issued by Medtronic Navigation, Inc..
- Company
- Medtronic Navigation, Inc.
- Recall Initiated
- November 28, 2023
- Posted
- January 10, 2024
- Recall Number
- Z-0699-2024
- Quantity
- 4,224
- Firm Location
- Lafayette, CO
- Official Source
- View on FDA website ↗
Reason for Recall
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Oman, Portugal, Spain, Switzerland, United Kingdom.
Lot / Code Info
UDI: 00613994247865 and 00643169105669(Japan)/ Serial Numbers: 2023041143, 2023051122, 2023051457, 2023051458, 2023060918, 2023051459, 2022030438, 2023041136, 2023041141
Root Cause
Device Design
Action Taken
On December 6, 2023, Medtronic Navigation, Inc. issued a Urgent: Medical Device Recall Notification via: UPS . Medtronic ask consignees to take the following actions: 1. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below. 2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form 3. Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at neuro.quality@medtronic.com. If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information. 4. This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.