RecallDepth

Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777

This recall is currently active, issued October 13, 2023. It was issued by Cardinal Health 200, Llc.

Company
Cardinal Health 200, Llc
Recall Initiated
September 13, 2023
Posted
October 13, 2023
Recall Number
Z-0112-2024
Quantity
3.927,994 units
Firm Location
Waukegan, IL
Official Source
View on FDA website ↗

Reason for Recall

Syringe contains conical tip that is not compatible with certain needleless IV connector(s).

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY

Lot / Code Info

Model Number/Product Code: 1180100777. UDI/DI: 10192253034530 each, 20192253034537 box, 50192253034538 case. Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601.

Root Cause

Process change control

Action Taken

Cardinal Health notified consignees on about 09/13/2023 via letter titled "URGENT MEDICAL DEVICE PRODUCT CORRECTION." Customers were instructed to review inventory for the affected product code and lot, ensure all personnel are informed cease utilizing the affected product with needleless IV connectors, post a copy of the notification in the storeroom where the product is stored, and notify any customers if the product has been further distributed/forwarded. Additionally, customers were asked to complete and return the included acknowledgment form.

More recalls by Cardinal Health 200, Llc

View all recalls by this company →