Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of the heart chambers during ablation.

Company
Abbott
Recall Initiated
June 9, 2022
Posted
July 7, 2022
Recall Number
Z-1344-2022
Quantity
191 devices
Firm Location
Plymouth, MN

Reason for Recall

When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."

Distribution

US Nationwide distribution.

Lot / Code Info

UDI 01)05415067027665(17)230831(10)xxxxxxx, Batch Numbers 8409416, 8409675, 8410811

Root Cause

Device Design

Action Taken

The firm notified its consignees by hand delivered Medical Device Recall letter on 06/09/2022. The letter explained the problem and provided the following instructions: "Next Steps To help reduce risk, we recommend: " Do not use any remaining inventory from the affected lots listed in Appendix A. " Complete and return the accompanying Acknowledgement Form to Abbott. " Return all remaining unused affected devices to Abbott. Your Abbott representative can assist you in returning these devices and obtaining replacements. Please forward this notice to anyone within your organization who may need to be notified and maintain a record of this notice along with a copy of the completed Acknowledgement Form for your records."