ABL800 Flex Analyzer Model Numbers 393-800 and 393-801

Company
Radiometer Medical Aps Akandevej 21 Bronshoj Denmark
Recall Initiated
May 18, 2022
Recall Number
Z-1321-2022
Quantity
1489 analyzers

Reason for Recall

There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Bermuda, Canada, Italy, Norway, Austria, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, China, Croatia, Denmark, Iceland, Germany, Spain, Portugal, India, Brunei, Japan, United Kingdom, Ireland, Russia, Hungary, Australia, New Zealand, Tokelau, Switzerland, France, Czech Republic, Slovakia, South Africa, Botswana, Namibia, Poland, Turkey, Sweden, Finland, Estonia, Kazakhstan, Malaysia, Hong Kong, Mexico, South Korea, Singapore.

Lot / Code Info

All ABL800 Flex Analyzers containing software versions below V6.19 MR2; UDI/DI ABL810, 820, and 830, ABL805, ABL815, ABL825, ABL835: 05700693938004 ; UDI/DI ABL817, ABL827, ABL837: 05700693938011

Root Cause

Software design

Action Taken

On May 20, 2022, Radiometer issued a "Urgent: Medical Device Recall" notification to affected consignees via: Email and FedEx 2nd day Mail. Radiometer ask consignees to take the following actions: 1. For the affected micro measuring modes Radiometer kindly requests you to not use results for cNa+. The most efficient way to ensure this is to remove the parameter from the affected modes as per the procedures included within the letter. 2. Please complete the Recall Response Form and return to your Radiometer representative (recall@radiometeramerica.com).