Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.

Recall Initiated
June 13, 2022
Recall Number
Z-1307-2022
Quantity
53 devices (19 US; 34 OUS)
Firm Location
Leesburg, VA

Reason for Recall

Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.

Distribution

Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to Canada, EU, and Hong Kong.

Lot / Code Info

UDI-DI: 10888857326804 Lot PCMW

Root Cause

Under Investigation by firm

Action Taken

On June 13, 2022, the firm distributed Medical Device Recall letters to affected customers. Customers were informed that local Stryker Sales Representative or Stryker Branch or Agency has already removed any product still located in customer facilities at the time of the initiation of the removal. If you have any questions or concerns, please contact Regulatory Compliance at 201.749.8090.