RecallDepth

Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp puncture tool (0.040 Stylet or 17ga Needle) through the parenchyma.

Company
Argon Medical Devices, Inc
Recall Initiated
July 8, 2021
Recall Number
Z-2501-2021
Quantity
20 units
Firm Location
Athens, TX

Reason for Recall

As a result of design changes, sheaths have exhibited cracking/breaking at the tips.

Distribution

U.S. Nationwide distribution in the state of WA. O.U.S.: N/A

Lot / Code Info

Item Numbers: TPS005 (Stylet); TPS006 (Needle); UDI Code: 00886333225552 Lot Number: 11369609

Root Cause

Device Design

Action Taken

On 07/08/2021, the firm's sales representative/clinical Specialist contacted via telephone to inform the single customer that the Recalling Firm was pausing evaluations of the affected device and was requesting return of the device. The customer was also informed that the Recalling Firm was in the process of implementing a few changes to improve upon the product. The product was place on hold at the customer's facility and the Recalling Firm's sales representative was picked up and returned to the Recalling Firm on 07/21/2021.