Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR MRI SureScan. 2. Astra family: Astra XT DR MRI SureScan, Astra XT SR MRI SureScan.

Company
Medtronic Inc., Cardiac Rhythm And Heart Failure (Crhf)
Recall Initiated
April 6, 2021
Recall Number
Z-1525-2021
Quantity
575815 units
Firm Location
Mounds View, MN

Reason for Recall

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Distribution

Worldwide distribution.

Lot / Code Info

all codes

Root Cause

Software design

Action Taken

Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2- day UPS delivery of the consignee letter. Consignees were asked to confirm receipt of the notification. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Confirmation will be obtained per locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. A courtesy notification will be provided to physicians with affected devices.

More recalls by Medtronic Inc., Cardiac Rhythm And Heart Failure (Crhf)

View all recalls by this company →