Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

This recall is currently active, issued November 16, 2023. It was issued by Philips North America.

Company
Philips North America
Recall Initiated
October 9, 2023
Posted
November 16, 2023
Recall Number
Z-0319-2024
Quantity
1 unit
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).

Distribution

US Nationwide distribution in the state of MN.

Lot / Code Info

UDI-DI: 00884838103627; Lot Code: 1005

Root Cause

Under Investigation by firm

Action Taken

An URGENT Medical Device Correction notification letter dated 10/9/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users Below are short-term precautions to take until the permanent solution is installed. Remain vigilant and confirm gantry position during procedures. Review the Emergency Stop Procedure and Collision Prevention Guidelines provided in Section 6: Working with Specialized Exams of the IFU. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the software update (FCO 72800805) to resolve the issue. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).

More recalls by Philips North America

View all recalls by this company →