RecallDepth

CONTRA GRAY (soft) TURBO PLUS CUPS, REF 153112. Dental Prophylaxis Angles

Company
Young Dental Manufacturing I, Llc
Recall Initiated
October 6, 2023
Posted
November 9, 2023
Recall Number
Z-0285-2024
Quantity
12 units
Firm Location
Earth City, MO

Reason for Recall

one lot of Contra Gray (soft) Turbo Plus Cups Dental Prophylaxis Angles may contain Contra Petite Web LF Soft Purple Cup DPAs.

Distribution

US: IN, TN, MO

Lot / Code Info

UDI/DI 10302731631127, Lot Number 223466

Root Cause

Packaging process control

Action Taken

Young Dental Manufacturing CO 1, LLC issued a MEDICAL DEVICE RECALL notice to its consignees on 10/06/2023 by USPS certified mail with return receipt notification. The notice explained the issue and requested that the product be returned or destroyed. Distributors were directed to notify their customers. Replacement product will be provided directly from the recalling firm; refunds will be handled through the distributor and credit issued.