CONTRA GRAY (soft) TURBO PLUS CUPS, REF 153112. Dental Prophylaxis Angles
- Recall Initiated
- October 6, 2023
- Posted
- November 9, 2023
- Recall Number
- Z-0285-2024
- Quantity
- 12 units
- Firm Location
- Earth City, MO
Reason for Recall
one lot of Contra Gray (soft) Turbo Plus Cups Dental Prophylaxis Angles may contain Contra Petite Web LF Soft Purple Cup DPAs.
Distribution
US: IN, TN, MO
Lot / Code Info
UDI/DI 10302731631127, Lot Number 223466
Root Cause
Packaging process control
Action Taken
Young Dental Manufacturing CO 1, LLC issued a MEDICAL DEVICE RECALL notice to its consignees on 10/06/2023 by USPS certified mail with return receipt notification. The notice explained the issue and requested that the product be returned or destroyed. Distributors were directed to notify their customers. Replacement product will be provided directly from the recalling firm; refunds will be handled through the distributor and credit issued.