OT1000 Series Orthopedic Surgical Tables

Company
Steris Corporation
Recall Initiated
January 21, 2021
Terminated
November 10, 2021
Recall Number
Z-1149-2021
Quantity
272 tables total (Domestic: 227 tables, Foreign 45 tables)
Firm Location
Mentor, OH

Reason for Recall

On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

Distribution

Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.

Lot / Code Info

Model No.: OT1000, OT1100, OT1200

Root Cause

Process control

Action Taken

In January, 2021, Steris will issue an Urgent Medical Device Correction notice to customers via letter for the OT1000 Series Orthopedic Surgical Table. Steris has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.