RecallDepth

DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product Usage: The DIASTAT anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl- 70, Jo-1, dsDNA, histone and centromere antigens.The test may be used to screen out samples which are negative for all ANAs. Samples that give a positive test result should be further tested to identify the antigen-specific antibody or antibodies present. Quantitative/qualitative ELISA kits are available for the individual detection of Sm, Sm/RNP, Ro (SS-A), La (SS-B), dsDNA and centromere ANAs. Qualitative kits are available for the individual detection of Scl-70 and Jo-1. ANA detection represents one parameter in a multicriterion diagnostic process.

This recall has been terminated (originally issued October 15, 2015).

Company
Euro Diagnostica Ab
Recall Initiated
October 15, 2015
Terminated
March 29, 2019
Recall Number
Z-0929-2018
Quantity
1090
Firm Location
Malmo Sweden
Official Source
View on FDA website ↗

Reason for Recall

Complaint investigation concluded that although product quality requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the strong IgM conjugate, and the low reference control led to an increased risk of equivocal and/or false positive sample test results for the two FANA200 kit lots SS 1500 and SS 2009.

Distribution

Worldwide Distribution and US. Nationwide

Lot / Code Info

UDI: 07340058410486. Lot numbers: SS 1500 and SS 2009.

Root Cause

Nonconforming Material/Component

Action Taken

Euro Diagnostica sent a Customer notifications letter to affected customers via email on 10/15/2015. The letter identified the affected product problem and actions to be taken. For question contact Euro Diagnostica .