Morris Hysto Pack, part number PSS3198(A
This recall has been terminated (originally issued February 28, 2017).
- Recall Initiated
- January 11, 2017
- Posted
- February 28, 2017
- Terminated
- May 3, 2017
- Recall Number
- Z-1345-2017
- Quantity
- 156 kits
- Firm Location
- Billings, MT
- Official Source
- View on FDA website ↗
Reason for Recall
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
Distribution
Distributed in AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, OH, PA, TX, UT, VA, and WA.
Lot / Code Info
Lot number and expiration date 101121, 10/5/2019 101680, 10/27/2019 83882, 7/8/2018 87088, 8/16/2018 87496, 9/24/2018 88351, 11/11/2018 89549, 12/22/2018 94996, 1/30/2019 97899, 6/15/2019 98769, 6/25/2019
Root Cause
Nonconforming Material/Component
Action Taken
AMS notified their consignees by Email, Fax, and Phone on 11/11/2017. AMS sent the notification letter, dated January 11, 2017, and provided yellow stickers "Recall Notice" to customers. They are advised to over-label the affected AMS kits at their facilities. Customers were instructed to place sticker in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date on a kit. Customers with questions can call Vicki Davis at 321-527-7714 or send email to vdavis@alignedmedicalsolutions.com.