Boston Scientific Advanix Pancreatic Stent, Straight Leading Barb, 10F x 7cm UPN: M00536650

This recall has been terminated (originally issued November 25, 2019).

Company
Boston Scientific Corporation
Recall Initiated
October 21, 2019
Posted
November 25, 2019
Terminated
December 20, 2023
Recall Number
Z-0552-2020
Quantity
940 units
Firm Location
Marlborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life

Distribution

Nationwide Foreign: Austria Australia Belgium Bulgaria Croatia Czech Rep Denmark Finland France Germany Great Britain Italy Japan Pakistan Poland Russian Fed. Slovakia South Africa Spain Sweden Switzerland Thailand Turkey

Lot / Code Info

Lot Numbers: 21187998  21285271  21337147  21483841  21512595  21662812  21724126  21833852  21922988  22036149  22136471  22307745  22359735  22613955  22700891  23035017  23295961  23492660  23541961  24006978  24271828  24383293  24411526

Root Cause

Incorrect or no expiration date

Action Taken

Boston Scientific issued Notifications to consignees via Federal Express Priority mail on October 21, 2019. The notification instructs the account to check their inventory remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form.