(1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL, REF 040183000A, 25/box, Sterile, Rx; and (2) Argon Medical Devices Double Male LL Adapter, Special Adapter to Convert a Female Fitting to Male LL, REF 040184000A, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.

This recall has been terminated (originally issued October 16, 2018).

Company
Argon Medical Devices, Inc
Recall Initiated
October 16, 2018
Terminated
June 15, 2021
Recall Number
Z-0288-2020
Quantity
Total number in recall for all products in dist: 49,130 units
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.

Distribution

Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.

Lot / Code Info

(1) Double Female LL Adapter-special adapter to convert male fitting to a female luer lock - Lot numbers 11228606, 11227966, 11226655; and  (2) Double Male LL Adapter-special adapter to convert a female fitting to a male luer lock - Lot numbers 11227508, 11229885, 11230268, and 11231393.

Root Cause

Equipment maintenance

Action Taken

The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.