ultraView Universal Alkaline Phosphatase Red Detection Kit, Catalog Number 05269814001, model 760-501 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Company
- Ventana Medical Systems Inc
- Recall Initiated
- August 2, 2018
- Posted
- September 14, 2018
- Terminated
- November 22, 2021
- Recall Number
- Z-3004-2018
- Quantity
- 8211 units
- Firm Location
- Oro Valley, AZ
Reason for Recall
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Distribution
worldwide
Lot / Code Info
Lot Numbers: Y15071, Y18053, Y22469, UDI: 4015630972258
Root Cause
Under Investigation by firm
Action Taken
The firm disseminated their field safety notices by letter on 08/02/2018. The firm requested the destruction of the product.