RecallDepth

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Recall Initiated
March 7, 2018
Recall Number
Z-1420-2018

Reason for Recall

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Distribution

PA, and worldwide

Lot / Code Info

All Lots

Root Cause

Component design/selection

Action Taken

Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.