Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
- Recall Initiated
- March 7, 2018
- Recall Number
- Z-1420-2018
Reason for Recall
Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
Distribution
PA, and worldwide
Lot / Code Info
All Lots
Root Cause
Component design/selection
Action Taken
Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.