Artis zee/zeego, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
This recall has been terminated (originally issued November 17, 2016).
- Recall Initiated
- November 1, 2016
- Posted
- November 17, 2016
- Terminated
- June 21, 2017
- Recall Number
- Z-0602-2017
- Quantity
- 4,657 systems Worldwide
- Firm Location
- Malvern, PA
- Official Source
- View on FDA website ↗
Reason for Recall
The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.
Distribution
US Nationwide Distribution.
Lot / Code Info
Model Numbers: 10094135, 10094137, 10094139, 10094141, 10280959
Root Cause
Software in the Use Environment
Action Taken
Siemens mailed a Customer Safety Advisory Notice to affected customers to inform them of a corrective action they plan to perform to prevent a possible hazard to persons and equipment. The notice identified the associated risks, underlying issues, impact on system operation and actions to be taken.
More recalls by Siemens Medical Solutions Usa, Inc
NAEOTOM Alpha.Prime
Software applications:
syngo.CT Brain Quantification
syngo.CT Vessel Hyperdensi...
Feb 3, 2026
NAEOTOM Alpha.Pro
Software applications:
syngo.CT Brain Quantification
syngo.CT Vessel Hyperdensiti...
Feb 3, 2026
SOMATOM X.ceed
Software applications:
syngo.CT Brain Quantification
syngo.CT Vessel Hyperdensities...
Feb 3, 2026