Medtronic Intellis Spinal Cord Stimulator, Model Number A710, Clinical Programmer application version 1.3.80 - Product Usage: intended for use by clinicians in the programming of the following Medtronic neurostimulators (external and implantable) for pain therapy.

Company
Medtronic Neuromodulation
Recall Initiated
November 12, 2020
Recall Number
Z-0860-2021
Quantity
7960 downloads
Firm Location
Minneapolis, MN

Reason for Recall

A710 Intellis Clinician Application has a software issue that can result in the inability to program the Intellis implantable neurostimulation device.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, Latin America, Asia Pacific, Japan, China, and European regions.

Lot / Code Info

Software version 1.3.80

Root Cause

Software design (manufacturing process)

Action Taken

Medtronic notified the first consignee by letter on 11/12/2020 with a confirmation of their complaint and request to perform a correction to the device which put the device in a factory-reset state and restore functionality to allow programming and continued therapy. On 17 Dec 2020, Medtronic will begin issuing an Urgent Medical Device Correction notification to physicians regarding this validation error and providing instructions on the impact and appropriate mitigating actions if this issue presents.