Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter

This recall has been terminated (originally issued July 8, 2019).

Company
Medtronic Navigation, Inc.
Recall Initiated
July 8, 2019
Terminated
October 16, 2024
Recall Number
Z-1905-2020
Quantity
4654 units
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.

Distribution

US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK

Lot / Code Info

All kits with lot numbers from 0211041602 through 0217695790

Root Cause

Nonconforming Material/Component

Action Taken

The firm issued its first customer notification and customer response form on 7/8/2019. On 7/16/2019, a second notification was issued to allow customers to identify recalled product by lot number. On 10/3/2019, the firm issued a third customer notification to include three additional affected lots. The firm requested customers examine inventory for affected lot numbers and quarantine for return to the firm and to complete the attached customer response form.