RecallDepth

Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v17) bone densitometer

Company
Ge Healthcare, Llc
Recall Initiated
December 3, 2019
Posted
April 24, 2020
Recall Number
Z-1711-2020
Quantity
123 in total
Firm Location
Waukesha, WI

Reason for Recall

An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.

Distribution

US Nationwide Distribution

Lot / Code Info

all affected systems

Root Cause

Radiation Control for Health and Safety Act

Action Taken

GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

More recalls by Ge Healthcare, Llc

View all recalls by this company →