Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
This recall has been terminated (originally issued July 23, 2019).
- Recall Initiated
- July 23, 2019
- Terminated
- June 8, 2020
- Recall Number
- Z-0042-2020
- Quantity
- 27676 units
- Firm Location
- Indianapolis, IN
- Official Source
- View on FDA website ↗
Reason for Recall
Inaccurately Low Sample Results
Distribution
AK AL AR AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY
Lot / Code Info
All Lots
Root Cause
Under Investigation by firm
Action Taken
1. Consignees will be provided instruction to download the updated e-barcode either from the cobas link or from the cobas e-library when available. 2. If analyzers cobas link is not connected to the Roche network, a future version of the cobas e-library CD will contain the revised e-barcodes. Download the e-barcode when available. 3. If analyzers cobas link is not connected to the Roche network and they require the updated VANC3 application e-barcode before the cobas e-library CD is available, they will be instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336 to request a Field Application Specialist to provide the VANC3 updated application. 4. Calibrate and run quality controls (QC) before running samples. 5.Complete all sections of the enclosed fax form. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.