CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

This recall has been terminated (originally issued February 23, 2018).

Company
Civco Medical Instruments Co. Inc.
Recall Initiated
February 23, 2018
Terminated
October 6, 2020
Recall Number
Z-2888-2018
Quantity
177 boxes
Firm Location
Kalona, IA
Official Source
View on FDA website ↗

Reason for Recall

A cover failed leak testing during validation testing of new needle guides from a replacement tool.

Distribution

Worldwide Distribution - US Nationwide in the states of GA, KY, CA, NY , NJ, TX, IL, MA, RI, AZ, and IA and countries of Ukraine, Hong Kong, Italy, Netherlands, Switzerland, United Kingdom, Denmark, Germany, Belgium, Spain, Japan, China, and Canada

Lot / Code Info

Item: 639-026 - Biopsy Starter Kit for use with Esaote LA332 and SL332 transducers Lot #'s M206950, M664690, M897110, M119380, M605540, M706720, M570610, M543670, M851530, M983170, M358570. M358570, M427660. M526300, M526300, M734570, M912740, and M664690. Item 698-009: Biopsy Starter Kit for use with Zonare C9-3 transducers M427370, M103030, A010473, M103030, M771190, M427390, M293590, M427370, M105150, M089260, M515370, M293580, M888920, M427370, M220960, M558420, M560740, M909270, M096330, M388690, M558220, M695400, M085550, M917410, M419650, M879530, M964370, M093020, M723180, and A010459 Item: 698-016 Biposy Bracket for use with Zonare C9-3 transducers M567620 and M052860 Item 620-116 Biopsy Starter Kit for use with BK Medical 18L5 and 8870 transducers M080890, M559550, M425050, M056610, M092800, M041630, M186540, M122250, M335740, M043310, M499390, M483250, M109060, M275630, M912150, A023904, M457560, M051670, M091360, M895560, A015235, M202260, M869430, M056610, M027090, M308820, M831280, M547370, M253050, M083160, and M865470

Root Cause

Device Design

Action Taken

Telephone Script to be read: CIVCO is calling to notify our customers regarding a recall action being initiated. CIVCO has recently discovered a dimensional issue in our Infiniti bracket which may allow the needle guide to a puncture a cover. You are being notified, as your facility has been identified as having received product from one of the affected manufacturing lots. Advise on Action to be taken by Customer or Distributor: To return Infiniti needle guide, contact your Account Representative at 800-445-6741 or 319-248-6757. Alternately you may also use email at order@civco.com to schedule the return of your device. Additional information to obtain: Verification of Email contact to send notification letter and response letter.