AutoDELFIA Neonatal immunoreactive trypsin(ogen) (IRT) kit, intended for the quantitative determination of human immunoreactive trypsin(ogen) (IRT) in blood specimens. (UDI: (01)06438147298716(17)190430(10)656479); Product Number: B005-212

Company
Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6
Recall Initiated
September 25, 2017
Terminated
September 6, 2018
Recall Number
Z-1220-2018
Quantity
240
Firm Location
Turku Finland

Reason for Recall

The Quality Control Certificate for certain AutoDELFIA Neonatal IRT Kits is missing information required to assess the acceptability of measured control values which may result in a delay in reporting IRT results.

Distribution

Distributed to CT and TX.

Lot / Code Info

Lot Number: 656479, Batch Number: 1065647901, Expiration date: 2019-4-30

Root Cause

Process control

Action Taken

Actions to be taken: Inspect your inventory of AutoDELFIA Neonatal IRT kits for batch number 1065647901 and replace the Quality Control Certificate included in the kit packaging with the revised certificate attached to this letter. If you have already used the kit lot there is no need to retest the samples reported if the results of the controls included in the assay were within your established acceptance range. Complete and return the enclosed response form identifying the quantity of affected kits in your possession. Please inform others affected in your organization accordingly. To comply with regulatory requirements we require that you complete the enclosed Response Form and return it.

More recalls by Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6

View all recalls by this company →