INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Company
Canon Medical System, Usa, Inc.
Recall Initiated
June 22, 2023
Posted
August 11, 2023
Recall Number
Z-2392-2023
Quantity
12
Firm Location
Tustin, CA

Reason for Recall

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

Distribution

US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA

Lot / Code Info

UDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003

Root Cause

Nonconforming Material/Component

Action Taken

On 6/22/23 correction notices were emailed to customers who were advised of the following: If the table side control does not function then the procedure can continue using the operators console in the operator room, the switches on the back of the column and the switches on the side of the flat panel detector. The emergency stop should not be affected and will continue to function normally. Neither manual operation or auto positioning of the C-arm can be performed on the console during procedure, and the C-arm will not operate. A firm representative will contact you to schedule a time to replace the PWB on the console. Complete and return the customer reply form to RAffairs@us.medical.canon For questions, contact the InTouch team at intouchdispatch@us.medical.canon or your service representative at 800.521.1968

More recalls by Canon Medical System, Usa, Inc.

View all recalls by this company →