RecallDepth

The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patients age, previous biopsy, and digital rectal exam (DRE). Patients file (sample type, prior biopsy status, DRE status and age) are uploaded to the Specimen Processing Module (SPM). The results from the four assays are exported automatically to Lab Manager Application (B2 LIS) which automatically triggers the 4Kscore calculation when values of the four analytes are available. The 4Kscore test results are determined automatically by the validated laboratory information system algorithm calculation software.

Company
Bioreference Health, Llc
Recall Initiated
November 18, 2022
Posted
June 2, 2023
Recall Number
Z-1676-2023
Quantity
1 system (662 Patients test results affected)
Firm Location
Elmwood Park, NJ

Reason for Recall

SPM software anomalies that may lead to the generation of erroneous 4Kscore Test results

Distribution

US Nationwide distribution in the state of New Jersey.

Lot / Code Info

UDI: 00850038860004

Root Cause

Software Design Change

Action Taken

Customers were notified via telephone between 11/18/22 and 12/31/22. In addition, the following information was provided within the corrected test results sent to customers: The 4Kscore result has been corrected to reflect the history of a prostatic nodule and/or the prior negative biopsy result, which was inaccurately posted into the algorithm, due to a bioinformatics error.

More recalls by Bioreference Health, Llc