RecallDepth

INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

Company
Medtronic Sofamor Danek Usa Inc
Recall Initiated
February 16, 2023
Posted
April 19, 2023
Recall Number
Z-1413-2023
Quantity
13 units
Firm Location
Memphis, TN

Reason for Recall

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Lot / Code Info

GTIN 00643169554696, Lot/Batch numbers: H5806345

Root Cause

Under Investigation by firm

Action Taken

The firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit. Reference FA1315. The letter also includes specific patient recommendations.