INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.
- Company
- Medtronic Sofamor Danek Usa Inc
- Recall Initiated
- February 16, 2023
- Posted
- April 19, 2023
- Recall Number
- Z-1413-2023
- Quantity
- 13 units
- Firm Location
- Memphis, TN
Reason for Recall
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot / Code Info
GTIN 00643169554696, Lot/Batch numbers: H5806345
Root Cause
Under Investigation by firm
Action Taken
The firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit. Reference FA1315. The letter also includes specific patient recommendations.