OnGuard2 CSTD Bag Adaptor SP A Chemfort CSTD Product, Product code MG412143
- Recall Initiated
- March 8, 2022
- Recall Number
- Z-1057-2022
- Quantity
- 293,520
Reason for Recall
Disconnection of the OnGuard2 Bag Adaptor SP from infusion bags which may result in spillage of hazardous medications.
Distribution
Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Lot / Code Info
UDI: 37290115921435, Lots: UBE507, UBE508, UBE513, UBE516, UBE517, UBE509, UBE511, UBE512, UBE514, UBE515, UBE517, UBE518, UBE526, UBE530, UBE531, UBE534, UBE533, UBE535, UBE536
Root Cause
Equipment maintenance
Action Taken
The recall notice and acknowledgement form dated 08 March 2022 were sent under B. Braun cover letter on 03/10/2022 via Certified USPS mail. Distributors are to forward the recall notification to customers to extend the recall to the end user/customer level. Identify current inventory, cease use and quarantine product subject to recall. Do not destroy, return product and complete the Product Removal Acknowledgement form. If you are a direct consignee of B. Braun Medical Inc. a B. Braun customer service representative will contact you to arrange for return of product. If you have received this recall notice from a customer of B. Braun, please return product to your local distributor to ensure proper credit and replacement of material. OnGuard Spike Port Adaptor may be used as an alternative to replace the OnGuard2 Bag Adaptor SP. If you have questions contact Simplivia: Shay Shaham, Sunday to Friday, 08:00 to 19:00 (GMT+2) at the following phone number: 97246908826.